The Intoxilyzer 9000 (part 1 of 2)
Roll-Out
The Michigan State Police (MSP) initiated Intoxilyzer 9000 (Intoxilyzer) training for police officers statewide, commencing in 2023. In order to participate, officers were required to complete both preliminary breath test (PBT) operator certification and PBT calibration certification.
Then officers were required to complete a PowerPoint training before attending a two-hour hands-on training session in person with MSP.
In July 2023, MSP commenced the installation of the Intoxilyzer at each jail or police agency that possessed a DataMaster DMT (DataMaster) and replaced the existing machine. By mid-October 2023, the new device was expected to be installed in all areas of the state.
Technical Differences
The DataMaster utilizes a pair of filters to effectively differentiate ethanol (the sought-after consumable alcohol by the police) from other organic compounds.
The narrow bandpass optical filters, centered at 3.44 microns, are strategically designed to isolate the specific wavelengths of light absorbed by alcohol and effectively block out any other interfering wavelengths. These filters, placed in the optical path, ensure accurate measurement of the unknown substance by precisely filtering out unwanted light.
However, it is worth noting that methanol and acetone can also be absorbed at the 3.4 micron level, meaning that the device may potentially register acetone or methanol as ethanol. This issue becomes particularly critical for individuals with diabetes who are experiencing ketoacidosis, as they tend to have acetone present in their airways.
The Intoxilyzer has four filters, including one in the 9 micron level, where ethanol and acetone do not appear. This should mean the Intoxilyzer is better able to filter out acetone and methanol.
However, because CMI, Inc. (CMI), the company that sells the Intoxilyzer, will not sell an Intoxilyzer to any entity other than law enforcement, this principle cannot be tested.
The Intoxilyzer’s filters utilize pulsed infrared technology, effectively eliminating the need for chopper motors or mechanical filters in the analytical system. Instead, it employs a digitally controlled pulsed infrared source, eliminating the use of moving parts and thus enhancing the durability of the filter mechanism, ensuring long-lasting performance.
In addition, the DataMaster used a quartz internal standard to make sure the machine remained in calibration. Quartz is a stable material with a known absorption rate.
The DataMaster would move a thin quartz plate in front of the infrared beam and the machine would compare the absorption to the known absorption rate at its initial calibration. If it was within tolerance, the DataMaster would report, “internal standard verified.”
The Intoxilyzer has done away with the quartz internal standard in favor of an actual dry-gas test to verify that the machine remains in calibration.
While CMI touts the dry-gas test as a better internal standard, there are two issues that may make it not as good as a constant standard like the quartz plate.
First, the dry gas is supposed to be measured at 0.08, but the Intoxilyzer allows any result between 0.074 to 0.084 to verify the machine’s internal standard.
The target of 0.08, which is affected by changes in barometric pressure, is addressed by the dry-gas test through an automated adjustment process. This adjustment is made using an algorithm including a correction factor based on the barometer reading at the time of the test.
Related Articles
No Results Found
The page you requested could not be found. Try refining your search, or use the navigation above to locate the post.
More Posts
Security Systems – You should have one!!
For most... a security video system is a tool to protect themselves from liability, false accusations and identify activities around their home or business. For some,...
What’s required of you in a Michigan traffic stop?
July 24, 2015 Dashcam video released this week from a traffic stop shows how a confrontation between a Texas woman and the arresting officer escalated. Given the...
Detroit police make arrests at marijuana dispensary
The Detroit police raided and made arrests at marijuana dispensary in an article in the Detroit Free Press from July 14, 2015. Detroit police arrested two people and...
Michigan Medical Marihuana Patient Bill of Rights
On November 8th, 2008, by a majority of 63 percent, the citizens of the State of Michigan voted into law the constitutional initiative, Initiated Law 1 of 2008,...
Colorado Supreme Court: Employers can fire for off-duty pot use
The Colorado Supreme Court ruled Monday Jun 15, 2015, that Employers’ zero-tolerance drug policies trump Colorado’s medical marijuana laws. In a 6-0 decision,...
How a sex toy put spotlight on Michigan civil asset forfeiture laws targeted for reform
The headlines read... "How a sex toy put national spotlight on Michigan civil asset forfeiture laws targeted for reform" "State Legislators Reconsider Forfeiture...
Reform Today’s Forfeiture Laws
Everyday, I get calls to my office from medical marijuana patients and caregivers who have been raided or pulled over by police. Often times, these individuals are not...
KOMORN LAW NEWSLETTER ISSUE #1 May 2015
The Michigan Legal Advisor News Letters. Read the current newsletter from Michigan's #1 Medical Marijuana Defense Attorney Michael Komorn. KOMORN LAW NEWSLETTER ISSUE...
Attorney Michael Komorn Lectures Students at the U of M Law School
I wanted to give a huge thanks to University of Michigan Law School Professors Howard Bromberg, Mark Osbeck and Law School class. This past Thursday I had the honor of...
Jury Selection In Marihuana Cases
A jury trial is fundamental to our democratic system of government. Every American citizen should embrace this responsibility by participating, and ensure justice...